UPClytics
View all cases

UPC CFI 137/2025

Feb 21, 2025·EP3567356 · SNIFFING LEAK DETECTOR WITH SWITCHING VALVE AND BUFFER CHAMBER

Case details
Status
Case Closed (No infringement)
Action
Infringement
Category
Main infringement action
Parties
Claimants
Representatives
Division
Dusseldorf LD
Technology
  • Instruments
    Measurement
Language
English
First decided
Oct 6, 2026
How this compares
The court's record for this kind of case, counted in first-instance rulings. A figure in brackets is the number of rulings a rate rests on.
Infringement foundMedian days to first decision
All UPC53% (62/117)124
Dusseldorf LDLockedLocked
Judge Berenice ThomLockedLocked
Judge Ingo RinkenLockedLocked
Judge Marije KnijffLockedLocked
Judge Ulrike HolzLockedLocked

At the UPC: Inficon GmbH (2 cases)·Elitech Technology, Inc. (2 cases)

Decisions
  • 2026-10-06
    InfringementPublic

    Summary: The Düsseldorf LD dismissed Inficon's infringement action against Elitech over EP 3 567 356 (the METRUM gas detector does not realise feature 5, a buffer chamber permitting homogeneous mixing; the claimant's measurements were inconclusive, so no specific contesting under R. 171.2 RoP was needed) and dismissed Elitech's counterclaim for revocation, holding claim 1 novel over CC1 and inventive. Each side bears the costs of the claim it lost; the interim costs request was refused; values EUR 1m and EUR 1.5m, cost ceiling EUR 400,000.

    Show the legal issues (4)
    • No specific contesting required under R. 171.2 RoP where the claimant's measurements are unconvincing from the outset
    • Claim construction of 'buffer chamber' permitting homogeneous mixing
    • Inventive step per Meril v Edwards / Amgen v Sanofi: no pointer or motivation shown
    • No interim award of costs under R. 150.2 RoP where success is split
Document titleDatePublic
Hearing recording reference document2026-10-13Not public
Publication of decision2026-10-06Not public
Case closed2026-10-06Not public
Decision2026-10-06Not public
Decision2026-10-06Public
Receipt2026-08-26Not public
Cover sheet2026-08-26Not public
Document Claimant2026-08-25Not public
Oral procedure opened2026-08-25Not public
Other Documents2026-08-25Public
Accepted arguments
What the court agreed with — by party.
  • The attacked METRUM device does not have a buffer chamber in which incoming reference gas is homogeneously mixed with remaining gas (feature 5); the delay in Exhibit TW 11 is not attributable to the housing

    RespondentLegal basis: Claim 1, feature 5; R. 171.2 RoP
  • CC1's blank filter 436 is not a buffer chamber, and the defendants showed no pointer or motivation towards the claimed solution

    ClaimantLegal basis: Art. 54, 56 EPC; Art. 65 UPCA
Rejected arguments
What the court did not agree with — and why.
  • Claim 1 lacks novelty over CC1 and inventive step

    RespondentLegal basis: Art. 54, 56 EPC
  • The defendants' failure to produce their own measurements makes the claimant's measurements undisputed

    ClaimantLegal basis: R. 171.2 RoP